
Replacing Assemblies 6
16800 Series Portable Logic Analyzers Service Guide 119
To remove and replace the power supplies
600 watt power supply
The 600 watt power supply is on all models but may look
different than the one shown here.
1 Perform previous procedures:
• “To prepare the instrument for disassembly" on
page 116.
• “To remove and replace the cover" on page 118.
2 Press the release latches on the sides of the connector to
disconnect the top 600 watt power supply cable from the
module interface board (MIB).
3 Press the single release tab to disconnect the lower 600 watt
power supply cable from the MIB.
4 Press the release latches to disconnect the 600 watt line filter
cable.
WARNING
Hazardous voltages exist on the power supply. To avoid electrical
shock, disconnect the power from the instrument before
performing the following procedures. After disconnecting the
power, wait at least six minutes for the capacitors on the power
supply board to discharge before servicing the instrument.
Komentarze do niniejszej Instrukcji