34 16800 Series Portable Logic Analyzers Service Guide
3 Testing Performance
Equipment Required
The following equipment is required for the performance test
procedure.
Table 9 Equipment Required
Equipment Critical Specification Recommended Model/Part
Pulse Generator ≥ 260 MHz, two channels, differential
outputs, 150-180 ps rise/fall time (if
faster, use transition time converters)
Agilent or HP 8133A option 003
150 ps Transition Time Converter
(Qty 3)
Required if pulse generator’s rise time
is less than 150 ps. (Pulse generator
conditions: Voffset=1V, ∆V=500 mV.)
Required for 8133A opt. 003.
Agilent or HP 15435A
Oscilloscope bandwidth ≥ 1.5 GHz,
sampling rate ≥ 8 GSa/s
Agilent or HP 54845A/B or similar
SMA Coax Cable
(Qty 2)
>18 GHz bandwidth Agilent or HP 8120-4948
Flying Lead Probe Set with 5 ground
leads
(Qty 2)
no substitute Agilent or HP E5383A
Male BNC to Female SMA adapters
(Qty 2)
Cambridge Products CP-AD507
(see www.cambridgeproducts.com)
SMA/Flying Lead test connectors, (f)
SMA to (f) SMA to Flying Lead Probe
(Qty 3)
no substitute See “To Assemble the SMA/Flying
Lead Test Connectors" on page 28