Replacing Assemblies 6
16800 Series Portable Logic Analyzers Service Guide 147
To remove and replace the module interface board
1 Perform previous procedures:
• “To prepare the instrument for disassembly" on
page 116.
• “To remove and replace the cover" on page 118.
• “To remove and replace the front panel assembly" on
page 125.
• “To remove and replace a measurement card" on
page 134 (if applicable).
• “To remove and replace a pattern generator card" on
page 138 (if applicable).
• “To remove and replace the tray assembly" on
page 143.
2 Ensure that all cables are disconnected from the module
interface board.
3 Using a 13/16- inch deep- well nut driver, remove the 3
BNC nuts from the module interface board (MIB).
4 Remove the retaining screw from the MIB.
5 Slide the board toward the front of the chassis enough to
allow the 9 standoffs to clear their keyholes, then lift the
board out of the chassis.