To remove and replace the module interface board
1 Perform previous procedures:
“To prepare the instrument for disassembly" on
page 116.
“To remove and replace the cover" on page 118.
“To remove and replace the front panel assembly" on
page 125.
“To remove and replace a measurement card" on
page 134 (if applicable).
“To remove and replace a pattern generator card" on
page 138 (if applicable).
“To remove and replace the tray assembly" on
page 143.
2 Ensure that all cables are disconnected from the module
interface board.
3 Using a 13/16- inch deep- well nut driver, remove the 3
BNC nuts from the module interface board (MIB).
4 Remove the retaining screw from the MIB.
5 Slide the board toward the front of the chassis enough to
allow the 9 standoffs to clear their keyholes, then lift the
board out of the chassis.