40 16800 Series Portable Logic Analyzers Service Guide
3 Testing Performance
6 Connect the E5383A Flying Lead Probe Set’s bits 2 and
10 to the SMA/Flying Lead test connector’s pin strip
connector at the 8133A pulse generator’s Channel 2
OUTPUT
.
7 Connect the E5383A Flying Lead Probe Set’s bits 6 and
14 to the SMA/Flying Lead test connector’s pin strip
connector at the 8133A pulse generator’s Channel 2
OUTPUT.
Connect the 8133A Pulse Generator Output to the 54845A Oscilloscope
1 Attach Male BNC to Female SMA adapters to Channels 1
and 2 on the 54845A oscilloscope.
2 Attach one end of an SMA cable to the Male BNC to
Female SMA adapter on Channel 1 of the oscilloscope.
3 Attach the other end of the SMA cable to the SMA/Flying
Lead connector at the Channel 2 OUTPUT of the 8133A
pulse generator.
4 Attach one end of the other SMA cable to the Male BNC
to Female SMA adapter on Channel 2 of the oscilloscope.
5 Attach the other end of the SMA cable to the SMA/Flying
Lead connector at the Channel 2 OUTPUT
of the 8133A
pulse generator.
NOTE
For each Flying Lead Probe connection, be sure to use a black ground lead
(five are supplied with the E5383A Flying Lead Probe Set) and orient the
leads so that the ground leads are connected to the SMA/Flying Lead
connector’s ground pins!
If you don’t have the ground leads, you can push the probe body’s ground
socket directly onto a ground pin on the SMA/Flying Lead test connector.
However, this will bend and could break pins on the SMA/Flying Lead test
connector because the probe body spacing is greater than the SMA/Flying
Lead test connector pin spacing. It is better to use the black ground leads.
In any case, the probe ground must be connected for each channel.