3 Testing Performance
Table 11 Oscilloscope Setup
Setup: Channel 1 Setup: Ch. 1 Probe Setup: Channel 2 Setup: Ch. 2 Probe
On Attenuation: 1.00:1 On Attenuation: 1.00:1
Scale: 100 mV/div Units: Volts Scale: 100 mV/div Units: Volts
Offset: 1 V Attenuation Units: Ratio Offset: 1 V Attenuation Units: Ratio
Coupling: DC External Gain: (n/a) Coupling: DC External Gain: (n/a)
Input: 50 ohm Skew: (Set later. See
page 42)
Input: 50 ohm Skew: 0.0 seconds
External Offset: (n/a) External Offset: (n/a)
Setup: Channel 3 Setup: Channel 4
Off Off
Setup: Horizontal Setup: Trigger Setup: Acquisition Setup: Display
Scale: 2 ns Mode: Edge Sampling Mode: Equiv. Time Waveforms: Connect dots
Position: 725 ps Source: Channel 1 Memory Depth: Automatic Persistence: Minimum
Reference: Center Level: 1.00 V Averaging: Enabled Grid: On (and set intensity)
Delayed: not selected Edge: Rising Edge # of Averages: 16 Backlight Saver: as preferred
Sweep: Auto
Measure: Markers
Mode: Manual placement
All else: (n/a)