Troubleshooting 5
16800 Series Portable Logic Analyzers Service Guide 101
Set up the test equipment
1 If you have not already done so, do the procedure “To
run the self-tests" on page 88.
2 Set up the pulse generator.
a Set up the pulse generator according to the following
table:
Connect the test equipment
1 Using two 2 x 9 test connectors, connect the logic
analyzer to the pulse generator channel outputs.
a Connect the even- numbered channels to the pulse
generator Channel 2 OUTPUT.
b Connect the odd- numbered channels the pulse
generator Channel 2 OUTPUT
.
2 Connect Clk1 to the pulse generator Channel 2 OUTPUT.
3 Enable the pulse generator Channel 1 and Channel 2
outputs (LEDs off).
Table 14 Pulse Generator Setup for Testing Cables
Timebase Pulse Channel 2 Trigger Pulse Channel 1
Mode: Int Mode: Pulse ÷ 1 Disable (LED on) Doesn’t matter, not used in this test.
Freq: 40 MHz Delay: (not available in pulse mode)
Width: 4 ns
Ampl: 3 V
Offs: 1.5 V
Output: Enable (LED off)
Comp: Normal (LED off)
Limit: Off (LED off)
Output
: Enable (LED off)