Testing Performance 3
16800 Series Portable Logic Analyzers Service Guide 45
Configure the Logic Analyzer
1 Exit the Agilent Logic Analyzer application (from the
main menu, choose FileExit) and then restart the
application. This puts the logic analyzer into its initial
state.
2 Specify the threshold voltage and assign channels to the
bus name:
a From the main menu, select SetupBus/Signal... or
SetupMy Logic Analyzer-1Bus/Signal....
b In the Analyzer Setup dialog, choose the Threshold
button for Pod 1.
Click here