Agilent Technologies 8133A Instrukcja Obsługi Strona 130

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 129
130 16800 Series Portable Logic Analyzers Service Guide
6 Replacing Assemblies
To remove and replace the display assembly
Your logic analyzer may have either a contrast filter or a touch
screen depending on the option you ordered.
1 Perform previous procedures:
“To prepare the instrument for disassembly" on
page 116.
“To remove and replace the cover" on page 118.
“To remove and replace the front panel assembly" on
page 125.
“To remove and replace the backlight inverter
board" on page 126.
“To remove and replace the front panel bracket
assembly" on page 128.
2 Using a Torx T10 screwdriver, remove the 2 screws securing
the display assembly to the front panel bracket assembly.
3 Remove the alignment pins on the bracket from the holes on
LCD.
T-10 screws (2)
Front panel
assembly
bracket
Liquid crystal
display (LCD)
Touch screen
control board
assembly
Contrast filter
(option 102)
Alignment
pins
Alignment
holes
Touch screen
(option 103)
Przeglądanie stron 129
1 2 ... 125 126 127 128 129 130 131 132 133 134 135 ... 187 188

Komentarze do niniejszej Instrukcji

Brak uwag