Agilent Technologies 8133A Instrukcja Obsługi Strona 125

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 124
Replacing Assemblies 6
16800 Series Portable Logic Analyzers Service Guide 125
To remove and replace the front panel assembly
1 Perform previous procedures:
“To prepare the instrument for disassembly" on
page 116.
“To remove and replace the cover" on page 118.
2 Disconnect the following cables:
USB cables from the motherboard.
Display cable from the display board (use care when
disconnecting to avoid damage).
Optional touch screen cable from the touch screen
controller board.
Keyboard cable from the keypad board.
Front panel cable from the backlight inverter board.
3 Using a Torx T10 screwdriver, remove 6 screws that secure
the front panel to the chassis.
4 Remove the front panel assembly.
5 Reverse this procedure to install the front panel assembly.
T-10
screws (3)
USB
cables
T-10
screws (3)
Display
cable
Touch screen
cable
Front panel
cable
Keyboard
cable