128 16800 Series Portable Logic Analyzers Service Guide
6 Replacing Assemblies
To remove and replace the front panel bracket assembly
1 Perform previous procedures:
“To prepare the instrument for disassembly" on
page 116.
“To remove and replace the cover" on page 118.
“To remove and replace the front panel assembly" on
page 125.
2 Using a Torx T10 screwdriver, remove the 2 screws securing
the touch screen controller board to the front panel bracket
assembly.
3 Using a Torx T10 screwdriver, remove the 6 screws securing
the front panel bracket assembly to the front panel.
4 Remove the bracket assembly.
5 Reverse this procedure to replace the front panel bracket
assembly.
T-1 0
screws (8)
Front
panel
Front panel
bracket assembly