118 16800 Series Portable Logic Analyzers Service Guide
6 Replacing Assemblies
To remove and replace the cover
1 Perform previous procedures:
• “To prepare the instrument for disassembly" on
page 116.
2 Using a Torx T20 screwdriver, remove the handle assembly.
3 Using a Torx T10 screwdriver, remove 6 screws that secure
the cover to the chassis.
4 Slide the cover back to remove.
5 Reverse this procedure to replace the cover.
T-20 screws (2)
T-10 screws (6)
CAUTION
When reinstalling the handle assembly, ensure that the screws are
torqued to 2.372 Newton meters (21 inch pounds) so that they do
not work themselves loose.