Testing Performance 3
16800 Series Portable Logic Analyzers Service Guide 57
Test Pod 2 in 250 Mb/s Mode
1 Disconnect the E5383A Flying Lead Probe Set from Pod 1
and connect it to Pod 2 of the logic analyzer. Do not
remove the flying leads that are connected to CLK and
the data channels.
2 On the 8133A pulse generator, in the PULSE setup for
CHANNEL 2, press the COMP button to return the
outputs to normal.
3 Note that the signal on the oscilloscope has moved.
Change the oscilloscope’s horizontal position to 725 ps (or
as required) to center the measured pulse on the
oscilloscope display.
4 Verify the DC offset and adjust it if necessary. See
page 41.
5 Deskew the oscilloscope if necessary. See page 42.
6 Readjust the pulse width from the pulse generator as
measured on the oscilloscope. See page 43.
7 From the Agilent Logic Analyzer application’s main
menu, select Setup→Bus/Signal... or Setup→My Logic
Analyzer-1→Bus/Signal...
8 Scroll to the right and unassign all Pod 1 bits.
9 Ensure that the Pod 2 threshold is set to 1 volt (see
page 46).
10 Assign bits 2, 6, 10, and 14 of Pod 2.
11 Select the Sampling tab (at the top of the dialog). In the
State Options area, set clock Clk1 to Don’t Care.
12 Set Clk2 to Both Edges.