134 16800 Series Portable Logic Analyzers Service Guide
6 Replacing Assemblies
To remove and replace a measurement card
Your product may consist of 1 or 2 measurement cards.
1 Perform previous procedures:
“To prepare the instrument for disassembly" on
page 116.
2 Use a Torx T10 screwdriver to remove the 4 screws
securing the shroud to the frame.
3 Slide the shroud over the cables.
4 Use a gentle back- and-forth motion as you pull to
disconnect the card(s) from the connector on the module
interface board and then pull the card(s) out of the
frame.
5 Replace the measurement card(s).
T-10
screws
(4)
Shroud