Test the Remaining Pods in 250 Mb/s Mode
The logic analyzer master card’s Pod 1 clock (“Clk1” in the
Sampling setup dialog) is used for testing all Pods above 4 in
the 250 Mb/s mode. Therefore, two E5383A Flying Lead
Probe Sets are required to test the remaining pods.
Set up the second E5383A Flying Lead Probe Set
1 Disconnect the first E5383A Flying Lead Probe Set from
the pod it is currently connected to (Pod 4) and connect
it to the pod under test (Pod 5).
2 Remove the pod under test Flying Lead Probe Set’s CLK
lead from the SMA/Flying Lead test connectors. (Do not
remove the channel 2, 6, 10, and 14 leads from the
SMA/Flying Lead test connectors.)
3 Connect a second E5383A Flying Lead Probe Set to Pod 1
of the logic analyzer’s master card.
4 Connect the Pod 1 Flying Lead Probe Set’s CLK lead to
the SMA/Flying Lead test connector’s pin strip connector
at the 8133A pulse generator’s Channel 1 OUTPUT.
Test the remaining pods
1 Perform the normal and complement tests for each
remaining pod on the logic analyzer, changing the
connection to the pod, channel assignments, thresholds,
etc. as appropriate.
Test using the logic analyzer master card’s Pod 1 clock,
“Clk1”, for all Pods above 4.
Upon completion, the logic analyzer is completely tested
in the 250 Mb/s mode.