50 16800 Series Portable Logic Analyzers Service Guide
3 Testing Performance
The important point is that your eye finder display
should look similar to the picture below (although it may
be shifted left or right), and eye finder must be able to
place the blue bars in the narrow eye. (The example
below shows eye finder in the 250 Mb/s mode.)
To re-align a stray channel
If the blue bar for a particular bit does not appear in its eye
near the recommended starting position, then do the
following steps to realign the sampling position of the stray
channel.
In the following example, the sampling position of two
channels (My Bus 1[0] and My Bus 1[2]) must be realigned
with the sampling position of the other channels.
8 Using the mouse, drag the sample position (blue bar) of
the stray channel (channels “My Bus 1[0]” and “My Bus
1[2]” in the above example) so that it is in the same eye
as the other channels. The Suggested Position from eye
finder (green triangle) will also move to the new eye.
Repeat the above step for all remaining stray channels.
9 Click the Run button in the Thresholds and Sample