Testing Performance 3
16800 Series Portable Logic Analyzers Service Guide 39
Connect the Test Equipment
Connect the Logic Analyzer Pod to the 8133A Pulse Generator
1 Connect a Transition Time Converter (if required—see
Table 7, “Recommended Test Equipment,” on page 19) to
the 8133A pulse generator’s Channel 2 OUTPUT, Channel
2 OUTPUT
, and Channel 1 OUTPUT.
2 Connect the SMA/Flying Lead test connector (see “To
Assemble the SMA/Flying Lead Test Connectors" on
page 28) with the 50 ohm terminator to the Transition
Time Converter at the 8133A pulse generator Channel 1
OUTPUT. (If Transition Time Converters are not required,
connect the SMA/Flying Lead test connector directly to
the pulse generator output.)
3 Connect the two SMA/Flying Lead test connectors without
50 ohm terminators to the Transition Time Converters at
the 8133A pulse generator Channel 2 OUTPUT and
Channel 2 OUTPUT
. (If Transition Time Converters are
not required, connect the SMA/Flying Lead test
connectors directly to the pulse generator outputs.)
4 Connect an E5383A Flying Lead Probe Set to Pod 1 of the
logic analyzer.
5 Connect the E5383A Flying Lead Probe Set’s CLK lead to
the pin strip of the SMA/Flying Lead connector at the
8133A pulse generator’s Channel 1 OUTPUT.
ground lead
Komentarze do niniejszej Instrukcji