Agilent Technologies 2000 X-Series Instrukcja Obsługi Strona 699

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 930
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 698
:WMEMory<r> Commands 32
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 699
:WMEMory<r>:SKEW
(see page 790)
Command Syntax
:WMEMory<r>:SKEW <skew>
<r> ::= 1-2 in NR1 format
<skew> ::= time in seconds in NR3 format
The :WMEMory<r>:SKEW command sets the skew factor for the specified
reference waveform.
Query Syntax
:WMEMory<r>:SKEW?
The :WMEMory<r>:SKEW? query returns the current skew setting for the
selected reference waveform.
Return Format
<skew><NL>
<skew> ::= time in seconds in NR3 format
See Also Chapter 32, “:WMEMory<r> Commands,” starting on page 693
":WMEMory<r>:DISPlay" on page 696
":WMEMory<r>:YOFFset" on page 700
":WMEMory<r>:YRANge" on page 701
":WMEMory<r>:YSCale" on page 702
Przeglądanie stron 698
1 2 ... 694 695 696 697 698 699 700 701 702 703 704 ... 929 930

Komentarze do niniejszej Instrukcji

Brak uwag