Agilent Technologies 2000 X-Series Instrukcja Obsługi Strona 397

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 930
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 396
:MTESt Commands 21
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 397
:MTESt:RMODe:FACTion:MEASure
(see page 790)
Command Syntax
:MTESt:RMODe:FACTion:MEASure <on_off>
<on_off> ::= {{1 | ON} | {0 | OFF}}
The :MTESt:RMODe:FACTion:MEASure command sets measuring only mask
failures on or off.
When ON, measurements and measurement statistics run only on
waveforms that contain a mask violation; passing waveforms do not affect
measurements and measurement statistics.
This mode is not available when the acquisition mode is set to Averaging.
Query Syntax
:MTESt:RMODe:FACTion:MEASure?
The :MTESt:RMODe:FACTion:MEASure? query returns the current mask
failure measure setting.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also "Introduction to :MTESt Commands" on page 379
":MTESt:RMODe:FACTion:PRINt" on page 398
":MTESt:RMODe:FACTion:SAVE" on page 399
":MTESt:RMODe:FACTion:STOP" on page 400
Przeglądanie stron 396
1 2 ... 392 393 394 395 396 397 398 399 400 401 402 ... 929 930

Komentarze do niniejszej Instrukcji

Brak uwag