:TRIGger Commands 29
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 615
:TRIGger:GLITch:SOURce
(see page 790)
Command Syntax
:TRIGger:GLITch:SOURce <source>
<source> ::= {DIGital<d> | CHANnel<n>}
<n> ::= 1 to (# analog channels) in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
The :TRIGger:GLITch:SOURce command selects the channel that produces
the pulse width trigger.
Query Syntax
:TRIGger:GLITch:SOURce?
The :TRIGger:GLITch:SOURce? query returns the current pulse width
source. If all channels are off, the query returns "NONE".
Return Format
<source><NL>
See Also "Introduction to :TRIGger Commands" on page 589
":TRIGger:MODE" on page 598
":TRIGger:GLITch:LEVel" on page 611
":TRIGger:GLITch:POLarity" on page 612
":TRIGger:GLITch:QUALifier" on page 613
":TRIGger:GLITch:RANGe" on page 614
Example Code "Example Code" on page 606