:MTESt Commands 21
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 399
:MTESt:RMODe:FACTion:SAVE
(see page 790)
Command Syntax
:MTESt:RMODe:FACTion:SAVE <on_off>
<on_off> ::= {{1 | ON} | {0 | OFF}}
The :MTESt:RMODe:FACTion:SAVE command sets saving on mask failures
on or off.
See Chapter 24, “:SAVE Commands,” starting on page 425 for more
information on save options.
Query Syntax
:MTESt:RMODe:FACTion:SAVE?
The :MTESt:RMODe:FACTion:SAVE? query returns the current mask failure
save setting.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also "Introduction to :MTESt Commands" on page 379
":MTESt:RMODe:FACTion:MEASure" on page 397
":MTESt:RMODe:FACTion:PRINt" on page 398
":MTESt:RMODe:FACTion:STOP" on page 400
NOTE
Setting :MTESt:RMODe:FACTion:SAVE ON automatically sets
:MTESt:RMODe:FACTion:PRINt OFF.