:CALibrate Commands 9
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 211
:CALibrate:OUTPut
(see page 790)
Command Syntax
:CALibrate:OUTPut <signal>
<signal> ::= {TRIGgers | MASK | WAVEgen}
The CALibrate:OUTPut command sets the signal that is available on the
rear panel TRIG OUT BNC:
TRIGgers — pulse when a trigger event occurs.
MASK — signal from mask test indicating a failure.
WAVEgen — waveform generator sync output signal. This signal depends
on the :WGEN:FUNCtion setting:
Query Syntax
:CALibrate:OUTPut?
The :CALibrate:OUTPut query returns the current source of the TRIG OUT
BNC signal.
Return Format
<signal><NL>
<signal> ::= {TRIG | MASK | WAVE}
See Also "Introduction to :CALibrate Commands" on page 207
":WGEN:FUNCtion" on page 667
Waveform
Type
Sync Signal Characteristics
SINusoid,
SQUare,
RAMP,
PULSe
The Sync signal is a TTL positive pulse that occurs when the waveform rises
above zero volts (or the DC offset value).
DC, NOISe N/A