328 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
19 :MARKer Commands
:MARKer:Y2Position
(see page 790)
Command Syntax
:MARKer:Y2Position <position> [suffix]
<position> ::= Y2 cursor position in NR3 format
<suffix> ::= {mV | V | dB}
If the :MARKer:MODE is not currently set to WAVeform (see
":MARKer:MODE" on page 319), the :MARKer:Y1Position command:
Sets :MARKer:MODE to MANual.
Sets the Y2 cursor position to the specified value.
Y cursor units are set by the :MARKer:YUNits command.
When the :MARKer:MODE is set to WAVeform, Y positions cannot be set.
Query Syntax
:MARKer:Y2Position?
The :MARKer:Y2Position? query returns current Y2 cursor position. This is
functionally equivalent to the obsolete :MEASure:VSTOp command/query.
Return Format
<position><NL>
<position> ::= Y2 cursor position in NR3 format
See Also "Introduction to :MARKer Commands" on page 318
":MARKer:MODE" on page 319
":MARKer:X1Y1source" on page 321
":MARKer:X2Y2source" on page 323
":MARKer:Y1Position" on page 327
":MARKer:YUNits" on page 330
":MEASure:VSTOp" on page 736
NOTE
If the front-panel cursors are off or are set to Binary or Hex Mode, the marker position
values are not defined and an error is generated. Make sure to set :MARKer:MODE to
MANual or WAVeform to put the cursors in the front-panel Normal mode.