Agilent Technologies 2000 X-Series Instrukcja Obsługi Strona 395

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 930
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 394
:MTESt Commands 21
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 395
:MTESt:LOCK
(see page 790)
Command Syntax
:MTESt:LOCK <on_off>
<on_off> ::= {{1 | ON} | {0 | OFF}}
The :MTESt:LOCK command enables or disables the mask lock feature:
ON — Locks a mask to the SOURce. As the vertical or horizontal scaling
or position of the SOURce changes, the mask is redrawn accordingly.
OFF — The mask is static and does not move.
Query Syntax
:MTESt:LOCK?
The :MTESt:LOCK? query returns the current mask lock setting.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also "Introduction to :MTESt Commands" on page 379
":MTESt:SOURce" on page 409
Przeglądanie stron 394
1 2 ... 390 391 392 393 394 395 396 397 398 399 400 ... 929 930

Komentarze do niniejszej Instrukcji

Brak uwag