:TRIGger Commands 29
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 591
General :TRIGger Commands
Table 89 General :TRIGger Commands Summary
Command Query Options and Query Returns
:TRIGger:FORCe (see
page 592)
n/a n/a
:TRIGger:HFReject {{0
| OFF} | {1 | ON}}
(see page 593)
:TRIGger:HFReject?
(see page 593)
{0 | 1}
:TRIGger:HOLDoff
<holdoff_time> (see
page 594)
:TRIGger:HOLDoff?
(see page 594)
<holdoff_time> ::= 60 ns to 10 s
in NR3 format
:TRIGger:LEVel:ASETup
(see page 595)
n/a n/a
:TRIGger:LEVel:HIGH
<level>, <source>
(see page 596)
:TRIGger:LEVel:HIGH?
<source> (see
page 596)
<level> ::= .75 x full-scale
voltage from center screen in NR3
format.
<source> ::= CHANnel<n>
<n> ::= 1 to (# analog channels)
in NR1 format
:TRIGger:LEVel:LOW
<level>, <source>
(see page 597)
:TRIGger:LEVel:LOW?
<source> (see
page 597)
<level> ::= .75 x full-scale
voltage from center screen in NR3
format.
<source> ::= CHANnel<n>
<n> ::= 1 to (# analog channels)
in NR1 format
:TRIGger:MODE <mode>
(see page 598)
:TRIGger:MODE? (see
page 598)
<mode> ::= {EDGE | GLITch |
PATTern | TV}
<return_value> ::= {<mode> |
<none>}
<none> ::= query returns "NONE"
if the :TIMebase:MODE is ROLL or
XY
:TRIGger:NREJect {{0
| OFF} | {1 | ON}}
(see page 599)
:TRIGger:NREJect?
(see page 599)
{0 | 1}
:TRIGger:SWEep
<sweep> (see
page 600)
:TRIGger:SWEep? (see
page 600)
<sweep> ::= {AUTO | NORMal}
Przeglądanie stron 590