622 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
29 :TRIGger Commands
:TRIGger:TV:LINE
(see page 790)
Command Syntax
:TRIGger:TV:LINE <line_number>
<line_number> ::= integer in NR1 format
The :TRIGger:TV:LINE command allows triggering on a specific line of
video. The line number limits vary with the standard and mode, as shown
in the following table.
Query Syntax
:TRIGger:TV:LINE?
The :TRIGger:TV:LINE? query returns the current TV trigger line number
setting.
Return Format
<line_number><NL>
<line_number>::= integer in NR1 format
See Also "Introduction to :TRIGger Commands" on page 589
":TRIGger:TV:STANdard" on page 626
":TRIGger:TV:MODE" on page 623
Table 94 TV Trigger Line Number Limits
TV Standard Mode
LFIeld1 LFIeld2 LALTernate
NTSC 1 to 263 1 to 262 1 to 262
PAL 1 to 313 314 to 625 1 to 312
PAL-M 1 to 263 264 to 525 1 to 262
SECAM 1 to 313 314 to 625 1 to 312