324 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
19 :MARKer Commands
:MARKer:XDELta
(see page 790)
Query Syntax
:MARKer:XDELta?
The MARKer:XDELta? query returns the value difference between the
current X1 and X2 cursor positions.
Xdelta = (Value at X2 cursor) - (Value at X1 cursor)
X cursor units are set by the :MARKer:XUNits command.
Return Format
<value><NL>
<value> ::= difference value in NR3 format.
See Also "Introduction to :MARKer Commands" on page 318
":MARKer:MODE" on page 319
":MARKer:X1Position" on page 320
":MARKer:X2Position" on page 322
":MARKer:X1Y1source" on page 321
":MARKer:X2Y2source" on page 323
":MARKer:XUNits" on page 325
NOTE
If the front-panel cursors are off, the marker position values are not defined. Make sure to
set :MARKer:MODE to MANual or WAVeform to put the cursors in the front-panel Normal
mode.