596 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
29 :TRIGger Commands
:TRIGger:LEVel:HIGH
(see page 790)
Command Syntax
:TRIGger:LEVel:HIGH <level>, <source>
<level> ::= 0.75 x full-scale voltage from center screen in NR3 format
for internal triggers
<source> ::= CHANnel<n>
<n> ::= 1 to (# analog channels) in NR1 format
The :TRIGger:LEVel:HIGH command sets the high trigger voltage level
voltage for the specified source.
Query Syntax
:TRIGger:LEVel:HIGH? <source>
The :TRIGger:LEVel:HIGH? query returns the high trigger voltage level for
the specified source.
Return Format
<level><NL>
See Also "Introduction to :TRIGger Commands" on page 589
":TRIGger:LEVel:LOW" on page 597
":TRIGger[:EDGE]:SOURce" on page 606