220 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
10 :CHANnel<n> Commands
:CHANnel<n>:BWLimit
(see page 790)
Command Syntax
:CHANnel<n>:BWLimit <bwlimit>
<bwlimit> ::= {{1 | ON} | {0 | OFF}
<n> ::= 1 to (# analog channels) in NR1 format
The :CHANnel<n>:BWLimit command controls an internal low-pass filter.
When the filter is on, the bandwidth of the specified channel is limited to
approximately 25 MHz.
Query Syntax
:CHANnel<n>:BWLimit?
The :CHANnel<n>:BWLimit? query returns the current setting of the
low- pass filter.
Return Format
<bwlimit><NL>
<bwlimit> ::= {1 | 0}
See Also "Introduction to :CHANnel<n> Commands" on page 218