Agilent Technologies B1500A Podręcznik Użytkownika Strona 93

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 628
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 92
Agilent B1500 Users Guide, Edition 7 2- 45
Introduction
Specifications
Period Width+2 ms (when Width 100 ms)
Period Width+10 ms (when Width > 100 ms)
Resolution: 100 μs
Pulse setting accuracy (fixed measurement range), supplemental information
Width: 0.5 % + 50 μs
Period: 0.5 % + 100 μs
SMU Supplemental Information
Maximum allowable cable resistance (Kelvin connection)
Sense: 10 Ω
Force: 10 Ω ( 100 mA), 1.5 Ω ( > 100 mA)
Voltage s
ource output resistance (Force line, Non-Kelvin connection)
0.2 Ω (HPSMU)
0.3 Ω (MPS
MU, HRSMU)
Voltage measu
rement input resistance: 10
13
Ω
Current source output resistance: 10
13
Ω (at 1 nA range)
Current compliance setting accuracy (for opposite polarity)
For 1 pA to 10 nA ranges: Setting accuracy ± 12 % of range
For 100 nA to 1 A ranges: Setting accuracy ± 2.5 % of range
Maximum capacitive load
For 1 pA to 10 nA ranges: 1000 pF
For 100 nA to 10 mA ranges: 10 nF
For 100 mA and 1 A ranges: 100 μF
Maximum guard capacitance
900 pF (HPSMU, MPSMU, HRSMU)
660 pF (HRSMU with ASU)
Maximum shield capacitance
5000 pF (HPSMU, MPSMU, HRSMU)
3500 pF (HRSMU with ASU)
Przeglądanie stron 92
1 2 ... 88 89 90 91 92 93 94 95 96 97 98 ... 627 628

Komentarze do niniejszej Instrukcji

Brak uwag