Agilent B1500 Users Guide, Edition 7 5- 53
Classic Test Definition
CMU Range Setup Window
CMU Range Setup Window
This window is opened by clicking the Range button on the C-V Sweep
Measurement Setup screen, and is used to set the measurement ranging operation.
Unit Module (MFCMU) used for the measurement
Mode Ranging mode, AUTO (auto ranging) or FIXED (fixed range).
If Mode=AUTO, the
MFCMU automatically searches for and
measures at the range that provides the best resolution.
If Mode=FIXED, the MFCMU measures at the range specified
b
y the Range field.
Ra
nge Measurement range. This field is effective when Mode=FI
XED.
Available measurement ranges depend on the measurement
frequency setting. See Table 5-3.
Table 5-3 Available Measurement Ranges
Frequency Available measurement ranges
200 kHz 50 Ω, 100
Ω, 300 Ω, 1 kΩ, 3 kΩ, 10 kΩ, 3
0 kΩ, 100 kΩ, 300 kΩ
2 MHz 50 Ω, 100 Ω, 300 Ω, 1 kΩ, 3 kΩ, 10
kΩ, 30 kΩ
5 MHz 50 Ω, 100 Ω, 300 Ω, 1 kΩ, 3 kΩ