Agilent Technologies B1500A Podręcznik Użytkownika Strona 441

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 628
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 440
Agilent B1500 Users Guide, Edition 7 7- 59
Function Details
SMU Measurement Time
Overhead Time
The overhead time is the time required for range changing and so on. This time
depends on the measurement condition, and cannot be specified. Major elements of
the overhead time are:
Range changing time during measurement (when measurement ranging mode is
set to auto or
limited auto)
Range changing time at measurement start (when the measur
ement range is less
than the compliance value)
To Specify Source Output Time
You cannot specify the measurement time. However, in sweep measurement mode,
you can specify a source output time that includes the measurement time. To specify
the source output time, set the delay time and the step delay time. The step delay
time is defined as the time from the measurement trigger up to the trigger for
changing the source output value. See the figure below. The step delay time must
cover the measurement time to keep the source output time constant.
The EasyEXPERT does not support the step delay time. It is available only for the
GP
IB remote mode.
9VWDUW
+ROGWLPH
0HDVXUHPHQWWLPH
'HOD\WLPH7G
6WHSGHOD\WLPH7VG
RQO\IRU5HPRWHPRGH
7G 7VG
7G 7VG
6RXUFHRXWSXWWLPH
Przeglądanie stron 440
1 2 ... 436 437 438 439 440 441 442 443 444 445 446 ... 627 628

Komentarze do niniejszej Instrukcji

Brak uwag