Agilent B1500 Users Guide, Edition 7 10-13
Application Library and Utilities
QSCV Measurement Supplemental Data
QSCV Measurement Supplemental Data
This section provides the following supplemental data of the QSCV measurement.
You can perform the QSCV measurement by using the QSCV[2] application test.
“Maximum Measurement Value”
“Considering Measurement Accuracy”
NOTE To obtain stable QSCV measurement results, use two source monitor units (SMU).
Assign the current meter and the voltage sweep source to the individual SMU.
Maximum Measurement Value
NOTE The maximum measurement value is not the specifications but the reference data.
Even if the capacitance is within this range, measurement may not be completed due
to influences on the measurement path such as extension cables, etc. The maximum
measurement value will be smaller depending on the leakage current.
The maximum measurement value depends on the settings of the current
measurement range, the QSCV measurement voltage, and the integration time. See
Figure 10-1 to Figure 10-5. Each figure shows the characteristics of the capacitance
value vs. the QSCV measurement voltage by the integration time setting. Each line
in the graph specifies the maximum measurement value for each integration time
setting.
The graph data is applied to the measurement when the QSCV integration time is set
to the same value as the LEAK integration time.