Agilent Technologies B1500A Podręcznik Użytkownika Strona 105

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 628
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 104
Agilent B1500 Users Guide, Edition 7 2- 57
Introduction
Specifications
Table 2-29 Pulse range and pulse parameter
Item
Value
a
a. Unless otherwise stated, all specifications assume a 50 Ω termination.
Frequency range 0.1 Hz to 33 MHz
Pulse period Programmable range 30 ns to 10 s
Resolution 10 ns
Minimum 100 ns (Vamp 10 V)
Accuracy
± 1 % (± 0.01 %
b
)
b. Typical minimum. This is the supplemental information.
Pulse width Programmable range 10 ns to (period -10 ns)
Resolution 2.5 ns (Tr and Tf 8 μs)
10 ns (Tr or Tf > 8 μs)
Minimum 50 ns (25 ns, typical) at Vamp 10 V
Accuracy ± (3 % + 2 ns)
Transition time (leading time
Tr and trailing time Tf)
c
c. The time from 10 % to 90 % of Vamp which is the amplitude of output pulse.
Programmable range 8 ns to 400 ms
Resolution 2 ns (Tr and Tf 8 μs)
8 ns (Tr or Tf > 8 μs)
Minimum (typical) < 15 ns (Vamp 10 V)
Minimum 20 ns (Vamp 10 V)
30 ns (Vamp 20 V)
60 ns (Vamp > 20 V)
Accuracy -5 % to 5 % + 10 ns (Vamp 10 V)
-5 % to 5
% + 20 ns (Vamp 20 V)
Output relay switching time
d
d. The time it takes the open state relay to open or close.
Open/close < 100 μs
Przeglądanie stron 104
1 2 ... 100 101 102 103 104 105 106 107 108 109 110 ... 627 628

Komentarze do niniejszej Instrukcji

Brak uwag