Agilent Technologies B1500A Podręcznik Użytkownika Strona 395

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 628
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 394
Agilent B1500 Users Guide, Edition 7 7- 13
Function Details
I/V-t Sampling Measurement
I/V-t Sampling Measurement
For the I/V-t sampling measurements, source channels (SMU) force constant current
or voltage, and monitor channels (SMU) measure current or voltage changes at a
device under test (DUT) with a specified sampling interval.
Available Modules
SMUs and SPGUs can be used. SMU can be constant voltage source or constant
current source and perform voltage measurement or current measurement. SPGU
can be pulse voltage source. See “SPGU Module” on page 7-26 for using the SPGU.
NOTE To use the SPGU, the sampling interval must be 2 ms.
Operation Summary
I/V-t sampling measurement is executed as explained below. See Figure 7-9. In this
example, the Output Sequence is set to SEQUEN
TIAL. And the bias hold function
sets the bias to the Base value.
Figure 7-9 Operation Summary
In this figure, a and b in the formula o
f Time indicate the time from the start of one
point measurement to the start of integration by the A/D converter.
9ROWDJHRUFXUUHQW
6WDUWVVDPSOLQJ
0HDVXUHPHQWWLPH
&KDQQHORXWSXW
&KDQQHORXWSXW
6RXUFHYDOXH
6RXUFHYDOXH
9ROWDJHRUFXUUHQW
7LPH
0HDVXUHPHQW
+ROGWLPH
,QWHUYDO
7LPH RULJLQ
7LPH +ROGWLPHD
7LPH +ROGWLPH,QWHUYDOE
D
E
%DVHKROGWLPH
%DVHYDOXH
Przeglądanie stron 394
1 2 ... 390 391 392 393 394 395 396 397 398 399 400 ... 627 628

Komentarze do niniejszej Instrukcji

Brak uwag