9-48 Agilent B1500 User’s Guide, Edition 7
If You Have a Problem
Error Codes
Agilent FLEX Execution Error
When Agilent B1500 causes errors in Agilent FLEX command execution
environment, the B1500 returns the following error code and error message.
100 Undefined GPIB command.
Send the correct command.
102 Incorrect numeric data syntax.
Correct the data syntax.
103 Incorrect terminator position.
Correct the command syntax. The number of parameters will be
inco
rrect.
104 Incorrect se
rial data syntax.
120 Incorrect parameter value.
Correct the parameter value.
121 Channel number must be 1 to 10.
Correct the channel number. The channel number must be 1 to 10 for
A
gilent B
1500.
122 Number of channels must be corrected.
Check the MM, FL, CN, CL, IN, DZ, or RZ command, and correct the
number of channe
ls.
123 Compliance must be set correctly.
Incorrect compliance value was set. Set the compliance value correctly.
124 In
correct range value for this channel.
Check the range value available for the channel, and correct the range
value.
125 Search goal v
alue must be less than compliance value.
126 Pulse base and peak must be same polarity.
The polarity of the base and peak values must be the same in the PI
co
mmand. Also the po
larity of the base, start, and stop values must be
the same in the PWI command.
130 Start and stop must be same polarity.