Agilent B1500 Users Guide, Edition 7 5- 15
Classic Test Definition
I/V List Sweep
Miscellaneous
Variables
This area defines the variables for the time stamp data and the index data.
Time Stamp Name Variable name of the time stamp. Time stamp data indicates
the time at which st
ep measurement was started.
Index Name Variable name of the data index. Integer above 0 is returned.
Measurement Setup
This setup screen is used to set the source channel outputs, the measurement ranging
mode, the sweep abort condition, the measurement channel A/D converter, the
series resistor, the filter, and the source output/measurement wait time.
VAR1 This area defines the primary sweep source output.
Unit Source monitor unit (SMU) used for the VAR1 output
Source Click the grid button to open the Define vector data dialog box.
See “Define vector data” on page 5-18.
Compliance Compliance value. Enter the current limit value for the voltage
source, or th
e voltage limit value for the current source.
Pwr Comp Power compliance value (in W), or OFF
NOTE To define VAR1/VAR2 step outputs
To define these parameters, create a vector data by using the Define vector data
dialog box which is opened by clicking the grid button in the vector data entry field.
See Figure 5-1.
Available array size for the vector data is as follows (column × row).
VAR1 step output value: N
1
= 1 × 1001 maximum (initial setting: 1 × 1)
VAR2 step output value: N
2
= 1 × 1001 maximum (initial setting: 1 × 1)
Delay time: N
3
= 1 × 1001 maximum (initial setting: 1 × 1)
Then, the following formula must be satisfied.
1 N
1
× N
2
128128