Agilent B1500 Users Guide, Edition 7 5- 27
Classic Test Definition
C-V Sweep
Channel Setup
This setup screen is used to define the channel setups and the variables for the
channel output data, time stamp data, and index data.
Channel Definition This area defines the channel setups and the v
ariables for th
e channel output data.
Unit Module used for the source output or measurement
V Name Available for the MFCMU and the SMU used for voltage
output. V
ariable name of voltage output data
I Name Available for the SMU used for current output. Variable name
of curren
t output data.
Mode SMU output mode, V (voltage), I (current), or COMMON
(ground)
Add Adds
a row for the channel setup.
Delete Deletes the channel setup. To select the channel setup, use the
left radi
o button.
Up Moves the selected channel setup upward.
Down Moves the selected channel setup downward.
If multiple output channels are defined, the channels start output sequentially. The
o
rder is
top to bottom of the channels defined in this screen. However the last output
channel is always the MFCMU. The output stops in the opposite order.
Miscellaneous
Va
riabl
es
This area defines the variables for the time stamp data and the index data.
Time Stamp Name Variable name of the time stamp. Time stamp data gives the
time when step measurem
ent is started.
Index Name Variable name of the data index. Integer above 0 is returned.