Agilent Technologies B1500A Podręcznik Użytkownika Strona 407

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 628
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 406
Agilent B1500 Users Guide, Edition 7 7- 25
Function Details
C-V Sweep Measurement
The MFCMU will internally perform the error correction for the CMU cables and
CMU accessories that keep the four-terminal pair configuration.
For the measurements for evaluating DUT against a working s
tandard, perform the
open/short/load correction.
However, you have to consider that some parasitic remain in measurement path
even after perfo
rming corrections. To minimize the influence of parasitic, make the
signal path from the shield ends to the DUT as short as possible.
To perform the phase compensation and error correction, see “CMU Calibration” on
page 4-37.
Figure 7-14 Error Correction
KLJK
&08UHWXUQ
ORZ
9LUWXDOJURXQG
0XOWLIUHTXHQF\&08
9
+FXU
$
/FXU
/SRW
$XWREDODQFHEULGJH
6\VWHPJURXQG
9[
2XWHUVKLHOG
LQWHUQDO
H[WHUQDO2SHQ6KRUW/RDG
9[
'87
WHUPLQDOSDLUV
WHUPLQDO
(UURUFRUUHFWLRQ
&RQQHFWLRQ
(UURUHOHPHQWV
+SRW
Przeglądanie stron 406
1 2 ... 402 403 404 405 406 407 408 409 410 411 412 ... 627 628

Komentarze do niniejszej Instrukcji

Brak uwag