Agilent B1500 Users Guide, Edition 7 2- 19
Introduction
Measurement Units
Table 2-1 HPSMU Output Voltage Ranges and Resolutions
Table 2-2 HPSMU Output Current Ranges and Resolutions
Range Output Value
Setting
Resolut
ion
Maximum
Current
2 V 0 |V| 2 V 100 μV ± 1000 mA
20 V 0 |V| 20 V 1 mV ± 1000 mA
40 V 0 |V| 40 V 2 mV ± 500 mA
100 V 0 |V| 100 V 5 mV ± 125 mA
200 V 0 |V| 200 V 10 mV ± 50 mA
Range Output Value
Setting
Resolution
Maximum
Voltage
1 nA 0 |I| 1.15 nA 50 fA ± 200 V
10 nA 0 |I| 11.5 nA 500 fA ± 200 V
100 nA 0 |I| 115 nA 5 pA ± 200 V
1 μA 0 |I|
1.15 μA 50 pA ± 200 V
10 μA 0 |I| 11.5 μA 500 pA ± 200 V
100 μA 0 |I| 115 μA 5 nA ± 200 V
1 mA 0 |I| 1.15 mA 50 nA ± 200 V
10 mA 0 |I| 11.5 mA 500 nA ± 200 V
100 mA 0 |I| 50 mA 5 μA ± 200 V
50 mA < |I| 115 mA 5 μA ± 100 V
1 A 0 |I| 50 mA 50 μA ± 200 V
50 mA < |I| 125 mA 50 μA ± 100 V
125 mA < |I| 500 mA 50 μA ± 40 V
500 mA < |I|
1 A 50 μA ± 20 V