Agilent Technologies E5071C Podręcznik Użytkownika Strona 408

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 551
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 407
408 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:SENS{1-4}:CORR:COLL:THRU
Syntax :SENSe{[1]|2|3|4}:CORRection:COLLect[:ACQuire]:THRU <numeric 1>,<numeric 2>
Description For channel 1 (:SENS1) to channel 4 (:SENS4), measure the calibration data of the thru
standard from the stimulus port to the response port. (No query)
Parameters
For each parameter, you must specify a different port number. If you specify the same port
number for 2 or more parameters, an error occurs and the command is ignored.
Example of use
10 OUTPUT 717;":SENS1:CORR:COLL:THRU 1,2"
20 OUTPUT 717;"*OPC?"
30 ENTER 717;A
Equivalent key [Cal] - Calibrate - Response (Thru) - Thru
[Cal] - Calibrate - n-Port Cal - Transmission - Port m-n Thru
<numeric 1> <numeric 2>
Description Response port number Stimulus port number
Range 1 to 2 1 to 2
Resolution 1 1
Przeglądanie stron 407
1 2 ... 403 404 405 406 407 408 409 410 411 412 413 ... 550 551

Komentarze do niniejszej Instrukcji

Vikiehs 04 Jan 2024 | 13:20:06

urenrjrjkvnm