130 Chapter 8
Limit Test
Sample Program
1140 !
1150 ! Checking test results
1160 !
1170 OUTPUT @Agte507x;":STAT:QUES:LIM?"
1180 ENTER @Agte507x;Reg_val
1190 Ch1_judge=BIT(Reg_val,1)
1200 OUTPUT @Agte507x;":STAT:QUES:LIM:CHAN1?"
1210 ENTER @Agte507x;Reg_val
1220 Tr1_judge=BIT(Reg_val,1)
1230 Tr2_judge=BIT(Reg_val,2)
1240 !
1250 ! Displaying test results
1260 !
1270 IF Ch1_judge=0 THEN
1280 PRINT "## PASS! ##"
1290 ELSE
1300 PRINT "## FAIL! ##"
1310 IF Tr1_judge=0 THEN
1320 PRINT " Trace1(S21): PASS"
1330 ELSE
1340 PRINT " Trace1(S21): FAIL"
1350 !
1360 ! Reading and displaying frequency at failed points
1370 !
1380 OUTPUT @Agte507x;":CALC1:PAR1:SEL"
1390 OUTPUT @Agte507x;":CALC1:LIM:REP:POIN?"
1400 ENTER @Agte507x;Fail_point
1410 REDIM Fail_data(1:Fail_point)
1420 OUTPUT @Agte507x;":CALC1:LIM:REP?"
1430 ENTER @Agte507x;Fail_data(*)
1440 PRINT " Frequency:"
1450 FOR I=1 TO Fail_point
1460 PRINT USING "3X,MD.4DE";Fail_data(I)
1470 NEXT I
1480 END IF
1490 IF Tr2_judge=0 THEN
1500 PRINT " Trace2(S11): PASS"
1510 ELSE
1520 PRINT " Trace2(S11): FAIL"
1530 !
1540 ! Reading and displaying frequency at failed points
1550 !
1560 OUTPUT @Agte507x;":CALC1:PAR2:SEL"
1570 OUTPUT @Agte507x;":CALC1:LIM:REP:POIN?"
1580 ENTER @Agte507x;Fail_point
1590 REDIM Fail_data(1:Fail_point)
1600 OUTPUT @Agte507x;":CALC1:LIM:REP?"
1610 ENTER @Agte507x;Fail_data(*)
1620 PRINT " Frequency:"
1630 FOR I=1 TO Fail_point
1640 PRINT USING "3X,MD.4DE";Fail_data(I)
1650 NEXT I
1660 END IF
1670 END IF
1680 END