232 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:CALC{1-4}:LIM
Syntax :CALCulate{[1]|2|3|4}[:SELected]:LIMit[:STATe] {ON|OFF|1|0}
:CALCulate{[1]|2|3|4}[:SELected]:LIMit[:STATe]?
Description For the active trace of channel 1 (:CALC1) to channel 4 (:CALC4) (specified with the
:CALC{1-4}:PAR{1-4}:SEL command), turns ON/OFF the limit test function.
Parameters
Query response {1|0}<newline><^END>
Example of use
10 OUTPUT 717;":CALC1:LIM ON"
20 OUTPUT 717;":CALC1:LIM?"
30 ENTER 717;A
Related commands :CALC{1-4}:PAR{1-4}:SEL on page 286
:CALC{1-4}:LIM:DISP on page 235
:DISP:FSIG on page 313
Equivalent key
[Analysis] - Limit Test - Limit Test
Description
ON or 1 Turns ON the limit test.
OFF or 0 (preset value) Turns OFF the limit test.