Agilent Technologies E5071C Podręcznik Użytkownika Strona 224

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 551
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 223
224 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:CALC{1-4}:FUNC:DOM:STAR
Syntax :CALCulate{[1]|2|3|4}[:SELected]:FUNCtion:DOMain:STARt <numeric>
:CALCulate{[1]|2|3|4}[:SELected]:FUNCtion:DOMain:STARt?
Description For channel 1 (:CALC1) to channel 4 (:CALC4), sets the start value of the analysis range
of the :CALC{1-4}:FUNC:EXEC command.
When the trace coupling is off, the active trace is the target to be set.
Parameters
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":CALC1:FUNC:DOM:STAR 1.7E9"
20 OUTPUT 717;":CALC1:FUNC:DOM:STAR?"
30 ENTER 717;A
Related commands :CALC{1-4}:FUNC:EXEC on page 225
:CALC{1-4}:FUNC:DOM on page 222
:CALC{1-4}:FUNC:DOM:COUP on page 223
:CALC{1-4}:FUNC:DOM:STOP on page 225
Equivalent key No equivalent key is available on the front panel.
<numeric>
Description Start value of the analysis range
Preset value 0
Unit Hz (hertz), dBm or second
Przeglądanie stron 223
1 2 ... 219 220 221 222 223 224 225 226 227 228 229 ... 550 551

Komentarze do niniejszej Instrukcji

Vikiehs 04 Jan 2024 | 13:20:06

urenrjrjkvnm