214 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:CALC{1-4}:CORR:OFFS:PHAS
Syntax :CALCulate{[1]|2|3|4}[:SELected]:CORRection:OFFSet:PHASe <numeric>
:CALCulate{[1]|2|3|4}[:SELected]:CORRection:OFFSet:PHASe?
Description Sets the phase offset of the active trace of channel 1 (:CALC1) to channel 4 (:CALC4)
(specified with the :CALC{1-4}:PAR{1-4}:SEL command).
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":CALC1:CORR:OFFS:PHAS 2.5"
20 OUTPUT 717;":CALC1:CORR:OFFS:PHAS?"
30 ENTER 717;A
Related commands :CALC{1-4}:PAR{1-4}:SEL on page 286
Equivalent key
[Scale] - Phase Offset
<numeric>
Description Phase offset
Range -360 to 360
Preset value 0
Unit ° (degree)