Agilent Technologies E5071C Podręcznik Użytkownika Strona 229

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 551
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 228
Chapter 13 229
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:CALC{1-4}:FUNC:TARG
Syntax :CALCulate{[1]|2|3|4}[:SELected]:FUNCtion:TARGet <numeric>
:CALCulate{[1]|2|3|4}[:SELected]:FUNCtion:TARGet?
Description For the active trace of channel 1 (:CALC1) to channel 4 (:CALC4) (specified with the
:CALC{1-4}:PAR{1-4}:SEL command), selects the target value when performing the
target search with the :CALC{1-4}:FUNC:EXEC command.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":CALC1:FUNC:TARG -12.5"
20 OUTPUT 717;":CALC1:FUNC:TARG?"
30 ENTER 717;A
Related commands :CALC{1-4}:FUNC:EXEC on page 225
:CALC{1-4}:FUNC:TTR on page 230
:CALC{1-4}:FUNC:TYPE on page 231
:CALC{1-4}:PAR{1-4}:SEL on page 286
Equivalent key No equivalent key is available on the front panel.
<numeric>
Description Target value
Range -5E8 to 5E8
Preset value 0
Unit Varies depending on the data format as follows:
Logarithmic Magnitude: dB (decibel)
Phase, Expand Phase, Positive Phase: ° (degree)
Group Delay: s (second)
Others: No unit
Przeglądanie stron 228
1 2 ... 224 225 226 227 228 229 230 231 232 233 234 ... 550 551

Komentarze do niniejszej Instrukcji

Vikiehs 04 Jan 2024 | 13:20:06

urenrjrjkvnm