286 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:CALC{1-4}:PAR{1-4}:SEL
Syntax :CALCulate{[1]|2|3|4}:PARameter{[1]|2|3|4}:SELect
Description Sets trace 1 (:PAR1) to trace 4 (:PAR4) of channel 1 (:CALC1) to channel 4 (:CALC4) to
the active trace.
You can set only the displayed trace to the active trace. If you execute this command trying
to set a not displayed trace to the active trace, an error occurs and the command is ignored.
(No query)
Example of use
10 OUTPUT 717;":CALC1:PAR1:SEL"
Related commands :DISP:WIND{1-4}:ACT on page 318
Equivalent key
[Trace Prev] / [Trace Next]
:CALC{1-4}:RLIM
Syntax :CALCulate{[1]|2|3|4}[:SELected]:RLIMit[:STATe] {ON|OFF|1|0}
:CALCulate{[1]|2|3|4}[:SELected]:RLIMit[:STATe]?
Description For the active trace of channel 1 (:CALC1) to channel 4 (:CALC4) (specified with the
:CALC{1-4}:PAR{1-4}:SEL command), turns ON/OFF the ripple test function.
Parameters
Query response {1|0}<newline><^END>
Example of use
10 OUTPUT 717;":CALC1:RLIM ON"
20 OUTPUT 717;":CALC1:RLIM?"
30 ENTER 717;A
Related commands :CALC{1-4}:PAR{1-4}:SEL on page 286
:CALC{1-4}:RLIM:DATA on page 287
:CALC{1-4}:RLIM:DISP:LINE on page 288
:CALC{1-4}:RLIM:DISP:SEL on page 289
:CALC{1-4}:RLIM:DISP:VAL on page 290
:CALC{1-4}:RLIM:FAIL? on page 291
:CALC{1-4}:RLIM:REP? on page 292
Equivalent key
[Analysis] - Ripple Limit - Ripple Limit Test
Description
ON or 1 Turns ON the ripple test function.
OFF or 0 (preset value) Turns OFF the ripple test function.