Agilent Technologies E5071C Podręcznik Użytkownika Strona 209

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 551
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 208
Chapter 13 209
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:CALC{1-4}:BLIM:MIN
Syntax :CALCulate{[1]|2|3|4}[:SELected]:BLIMit:MINimum <numeric>
:CALCulate{[1]|2|3|4}[:SELected]:BLIMit:MINimum?
Description For channel 1 (:CALC1) to channel 4 (:CALC4) (specified with the
:CALC{1-4}:PAR{1-4}:SEL command), sets the lower limit value of the bandwidth test.
Parameters
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":CALC1:BLIM:MIN 1E4"
20 OUTPUT 717;":CALC1:BLIM:MIN?"
30 ENTER 717;A
Related commands :CALC{1-4}:PAR{1-4}:SEL on page 286
:CALC{1-4}:BLIM on page 203
:CALC{1-4}:BLIM:MAX on page 208
Equivalent key
[Analysis] - Bandwidth Limit - Min Bandwidth
<numeric>
Description Minimum bandwidth.
Range 0 to 1E12
Preset value 300 k
Unit Hz (hertz), dB or second