Agilent Technologies E5071C Podręcznik Użytkownika Strona 283

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 551
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 282
Chapter 13 283
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:CALC{1-4}:MST:DATA?
Syntax :CALCulate{[1]|2|3|4}[:SELected]:MSTatistics:DATA?
Description For the active trace of channel 1 (:CALC1) to channel 4 (:CALC4) (specified with the
:CALC{1-4}:PAR{1-4}:SEL command), reads out the statistics values (the mean vale, the
standard deviation, and the difference between the maximum value and the minimum
value) display. (Query only)
Query response {numeric 1},{numeric 2},{numeric 3}<newline><^END>
Example of use
10 OUTPUT 717;":CALC1:MST:DATA?"
20 ENTER 717;A,B,C
Related commands :CALC{1-4}:MST on page 282
:CALC{1-4}:PAR{1-4}:SEL on page 286
Equivalent key No equivalent key is available on the front panel.
Description
{numeric 1} Mean value
{numeric 2} Standard deviation
{numeric 3} Difference between the maximum value and the minimum value
(Peak to Peak)