340 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:INIT{1-4}
Syntax :INITiate{[1]|2|3|4}[:IMMediate]
Description Changes the state of each channel of channel 1 (:INIT1) to channel 4 (:INIT4) to the startup
state in the trigger system.
When this command is executed for a channel in the idle state, it goes into the initiate state
immediately. Then, after measurement is executed once, it goes back to the idle state.
If this command is executed for a channel that is not in the idle state or for which the
continuous initiation mode is set to ON (ON is specified with the :INIT{1-4}:CONT
command), an error occurs and the command is ignored.
For details about the trigger system, refer to “Trigger system” on page 80. (No query)
Example of use
10 OUTPUT 717;":INIT1"
Related commands :INIT{1-4}:CONT on page 341
Equivalent key
[Trigger] - Single