1100 Series DAD and MWD Reference Manual 97
Repairing the Detector 4
Cleaning the Detector
The detector case should be kept clean. Cleaning should be done with a soft
cloth slightly dampened with water or a solution of water and mild detergent.
Do not use an excessively damp cloth allowing liquid to drip into the detector.
CAUTION
Electronic boards and components are sensitive to electronic discharge (ESD). In order
to prevent damage always use an ESD protection (for example, the ESD wrist strap
from the accessory kit) when handling electronic boards and components (see “Using
the ESD Strap” on page 98).
WARNING
Eye damage may result from directly viewing the light produced by the deuterium
lamp used in this product. Always turn the deuterium lamp off before removing it.
WARNING
Do not let liquid drip into the detector. It could cause shock hazard and it could
damage the detector.